🧪 Test Procedure
Please read the operating instructions carefully before starting the experiment. Allow the test card and test solution to return to room temperature before use. Remove the reagent container and test strip bag from the original packaging. After opening, take out the required number of microwell reagents and test cards, and label them. Use within 60 min. Using a pipette, add 200 μL of test sample solution into the microwell, mix thoroughly by slowly pipetting to ensure full mixing with the microwell reagent. Incubate at room temperature for 5 min, then using a pipette, vertically add approximately 100 μL of the mixed solution into the sample well (S well), or use the disposable dropper provided in the kit to slowly and vertically add 3 drops of the mixed solution into the sample well (S well). Start timing when the liquid begins to flow. Allow to react for 5 min, then immediately interpret the result according to the schematic diagram; results read at other times are invalid.