Home / Rapid Test Reagents / SulfonamidesRapid Test Strip

SulfonamidesRapid Test Strip

rapid testReagents · LFG00801 · Yetta Bio

SulfonamidesRapid Test Strip - rapid testReagents - Yetta Bio

SulfonamidesRapid Test Strip

Rapid Test Reagents · Food Safety Rapid Testing

LFG00801📦 10T📝 Colloidal Gold Strip · LOD0.1mg/kg · For livestock/poultry tissue, urine · 6 samples≤20 min

Colloidal Gold Strip · LOD0.1mg/kg · For livestock/poultry tissue, urine · 6 samples≤20 min

← Rapid Test Reagents List
📦
Spec
10T
🏷️
SKU
LFG00801
🎯
LOD
0.1 mg/kg
GB 31650-2019
🔬
Method
Colloidal Gold
🧪
Sample Type
livestock/poultry tissue, urine
⏱️
Test Time
6 samples≤20 min
Performance
Colloidal Gold · 6 samples≤20 min · For livestock/poultry tissue, urine · LOD0.1 mg/kg · Shelf Life: 12 months
🌡️
Storage
Store at 2–30°C away from light
📅
Shelf Life
12 months
📜
Regulatory Basis
GB 31650-2019
KJ 201705
🔧Required Equipment & Consumables
🔬Equipment
(1) homogenizer, (2) balance (resolution 0.01 g) , (3) vortex mixer, (4) Centrifuge , (5) Pipette Set (20-200µL, 100-1000µL) , (6) Colloidal Gold Analyzer (Optional)
📦Consumables
15 mL Centrifuge Tube
📋Test Procedure Guide
1
📝 Sample Preparation
Before testing,SampleshouldUsehomogenizer, SampleMash into Meat Paste. AccurateWeigh 1.00±0.05 gTissueSample至15 mL Centrifuge Tube, add 5 mLSample Extractant, Vortex2minSufficientlyMix well. 4000r/minroom temperatureCentrifuge5min. 取upper layer相forClearPart/Portionassample solution (Part/PortionSampleoils & fatstoo多, upper layerHas a LayerWhiteSuspension TurbidSolution, in this caseTake the Intermediate PhaseforClearPart/Portion) .
2
🧪 Test Procedure
Before the experiment, carefully read the operating instructions. Before use, allow the test card and sample solution to equilibrate to room temperature (20–25°C). Remove the test card from the original packaging, take out the required number of test cards, label them, and use immediately. Using a pipette, aspirate 100 μL of sample solution and vertically add it into the sample well (S well), or use the disposable dropper provided in the kit to aspirate sample solution and slowly add 3 drops vertically into the sample well (S well). Start timing when liquid begins to flow. React for 5 min, then immediately interpret the result according to the result diagram or read the test result using a colloidal gold reader. Interpretations at other times are invalid.
Purchase Inquiry