🧪 Test Procedure
Before the experiment, please read the operating instructions carefully. Equilibrate the test cassette and the sample to be tested to room temperature before use. Remove the reagent barrel and test strip bag from the original packaging. Take out the required number of microwell reagents and test cassettes, and mark them. Use within 60 min. Using a micropipette, aspirate 200 μL of the sample to be tested into the microwell, slowly aspirate and dispense to mix thoroughly with the microwell reagent. Incubate at room temperature for 5 min. Then, using a micropipette, aspirate approximately 100 μL of the mixed solution and add it vertically into the sample well (S well), or use the disposable dropper provided in the kit to slowly add 3 drops vertically into the sample well (S well). Start timing when the liquid flows. React for 5 min, then immediately interpret the result according to the schematic diagram. Results interpreted at other times are invalid.