Home / Rapid Test Reagents / Fumonisin Fluorescence Quantitative Card

Fumonisin Fluorescence Quantitative Card

rapid testReagents · LFF00604 · Yetta Bio

Fumonisin Fluorescence Quantitative Card - rapid testReagents - Yetta Bio

Fumonisin Fluorescence Quantitative Card

Rapid Test Reagents · Food Safety Rapid Testing

LFF00604📦 40T📝 Fluorescence Quantitative Card · 6 samples≤30 min · For grains, feed & processed products · CV<3%

Fluorescence Quantitative Card · 6 samples≤30 min · For grains, feed & processed products · CV<3%

← Rapid Test Reagents List
📦
Spec
40T
🏷️
SKU
LFF00604
🎯
LOD
1000 μg/kg
GB 2761-2017
🔬
Method
Fluorescence Quantitative
🧪
Sample Type
grains, feed & processed products
⏱️
Test Time
6 samples≤30 min
Performance
Fluorescence Quantitative · 6 samples≤30 min · For grains, feed & processed products · LOD1000 μg/kg · Shelf Life: 12 months
📅
Shelf Life
12 months
🔧Required Equipment & Consumables
🔬Equipment
(1) Pipette Set (20–200 μL, 100–1000 μL) , (2) balance: (0.01 g) , (3) vortex mixer, (4) Sample Grinder, (5) 500 mL Graduated Cylinder , (6) sieve, (7) Centrifuge (Min. Speed >4000r/ min) , (8) Constant Temp.Incubator , (9) Fluorescence Reader
🧪Reagents
(1) absolute ethanol (AR级) , (2) deionized water
📦Consumables
(1) 50 mL Centrifuge Tube, (2) 2 mL Centrifuge Tube
📋Test Procedure Guide
1
Sample Prep
-3.which得supernatant100ul, Add 200 μL 50%ethanol-water extraction solution中Mix well, fromTake from 100μL Add 500μL samplediluent中, Mix well后sample applicationDetection, DetectionResultManual乘3 即asmost终concentration.
2
Detection
Turn on the Thermostatic Incubator; ready to use once temperature reaches 37°C; Before the experiment, please read this manual and the fluorescence reader operating instructions carefully. Before use, allow the test cassette and the sample solution to equilibrate to room temperature. Remove the test cassette from the original packaging, mark it, and use immediately. Using a pipette, add 100 μL of the sample solution vertically into the sample well. Place on an incubator at 38°C and incubate for 8 minutes. Insert the test cassette (sample end) into the fluorescence reader to read the result.
Purchase Inquiry