🧪 Test Procedure
Before the experiment, carefully read the operating instructions. Before use, allow the test card and the sample to reach room temperature (20–25°C). Remove the test card from its original packaging, open it, label it, and use immediately. Using a pipette, aspirate 200 μL of the sample solution from the previous sample pretreatment step into a gold-labeled microwell. Gently pipette up and down for 30 seconds to completely dissolve the red substance at the bottom of the gold-labeled microwell, and let it stand horizontally. After a reaction wait of 3 minutes, use a pipette to aspirate the red solution in the gold-labeled microwell. Gently pipette up and down for 10 seconds, then aspirate 100 μL of the mixed solution and drip it into the sample well (S) of the test card. Start timing after adding the sample. Alternatively, use the disposable dropper provided in the kit to aspirate the mixed solution and slowly add 3 drops vertically into the sample well (S well). Start timing when the liquid flows. React for 5-8 min. Interpret the result according to the schematic diagram. Results at other times are invalid.